Power: 100 to 240 VAC, 50 or 60 Hz, nominal;
Attachable Modules: up to 8;
Display: 4.25″ x 2.25″ LCD graphics, LED backlight, adjustable contrast and brightness;
Automatic Input Sequencing: 1 to 96 channels;
Communications: RS-232;
Non-volatile Memory: channel sequence, probe coefficients;
Minimum Sample Time: 2 seconds.
Resistance Modules | ||||||||
. |
Input Channels
|
Resistance Range
|
Basic Resistance Accuracy
|
Resistance Resolution
|
Temp.
Range |
Equivalent Temperature Accuracy†
|
Temp.
Resolution |
Excitation Current
|
SPRT Module 2560 |
2
|
0Ω to 400Ω
|
±20 ppm of reading (0.0005Ω at 25Ω, 0.002Ω at 100Ω)
|
0.0001Ω
|
–260°C to 962°C
|
±0.005°C at 0°C ±0.007°C at 100°C
|
0.0001°C
|
1.0 mA,
1.4 mA |
High-Temp PRT Module 2561 |
2
|
0Ω to 25Ω
|
±50 ppm of reading (0.00013Ω at 2.5Ω)
|
0.00001Ω
|
0°C to 1200°C
|
±0.013°C at 0°C ±0.018°C at 100°C
|
0.001°C
|
3.0 mA,
5.0 mA |
PRT Scanner 2562 |
8
|
0Ω to 400Ω
|
±40 ppm of reading (0.004Ω at 100Ω)
|
0.0001Ω
|
–200°C to 850°C
|
±0.01°C at 0°C ±0.014°C at 100°C
|
0.0001°C
|
1.0 mA,
1.4 mA |
Standards Thermistor Module 2563 |
2
|
0Ω to 1 MΩ
|
±50 ppm of reading (0.5Ω at 10 KΩ)
|
0.1Ω
|
–60°C to 260°C
|
±0.0013°C at 0°C ±0.0015°C at 75°C
|
0.0001°C
|
2 µA,
10 µA |
Thermistor Scanner 2564 |
8
|
0Ω to 1 MΩ
|
±100 ppm of reading (1Ω at 10 KΩ)
|
0.1Ω
|
–60°C to 260°C
|
±0.0025°C at 0°C ±0.003°C at 75°C
|
0.0001°C
|
2 µA,
10 µA |
1000Ω PRT Module 2567 |
2
|
0Ω to 4 KΩ
|
±25 ppm of reading (0.025 at 1 KΩ)
|
0.001Ω
|
–260°C to 962°C
|
±0.006°C at 0°C ±0.009°C at 100°C
|
0.0001°C
|
0.1 mA,
0.05 mA |
1000Ω PRT Scanner 2568 |
8
|
0Ω to 4 KΩ
|
±40 ppm of reading (0.04 at 1 KΩ)
|
0.001Ω
|
–200°C to 850°C
|
±0.01°C at 0°C ±0.014°C at 100°C
|
0.0001°C
|
0.1 mA,
0.05 mA |
Thermocouple Modules | |||||||
. |
Input Channels
|
Millivolt Range
|
Millivolt Accuracy
|
Millivolt Resolution
|
Temperature Accuracy,† Ext. CJC
|
Temperature Accuracy,† Int. CJC
|
Temperature Resolution
|
Precision Thermocouple Module 2565 |
2
|
–10 to 100 mV
|
±0.002 mV
|
0.0001 mV
|
±0.05°C
|
±0.1°C
|
0.001°C
|
Thermocouple Scanner 2566 |
12
|
–10 to 100 mV
|
±0.004 mV
|
0.0001 mV
|
±0.1°C
|
±0.3°C
|
0.001°C
|
†Temperature accuracy depends on probe type and temperature |